| Max Capacity (grams) | Display | Draft Shield | SKU | Price |
|---|---|---|---|---|
| 120 g | MCE | Manual Draft Shield | SAR:MCE124SU-S00 |
$5,462
|
| 120 g | MCA | Manual Draft Shield | SAR:MCA124SU-S00 |
$8,117
|
| 120 g | MCA | Automatic Draft Shield with Ionizer | SAR:MCA124SI-S00 |
$11,331
|
| 220 g | MCE | Manual Draft Shield | SAR:MCE224SU-S00 |
$6,581
|
| 320 g | MCE | Manual Draft Shield | SAR:MCE324SU-S00 |
$7,349
|
| 320 g | MCA | Manual Draft Shield | SAR:MCA324SU-S00 |
$10,004
|
| 320 g | MCA | Automatic Draft Shield with Ionizer | SAR:MCA324SI-S00 |
$13,218
|
The modular weighing system Cubis® II can be individually configured from different components. The combinability of the display unit, weighing module, draft shield, software packages for various applications and functions and a comprehensive range of accessories allows the individual adaptation of the Cubis® II balance to all weighing tasks.
The Cubis® II range of premium laboratory balances with a maximum load between 2.1 g and 70 kg and a readability >between 0.1 μg to 1 g provide the ideal model for every application.
Features:
Display and Control Unit Options:
Draft Shield Options:
| Capacity | 120 g, 220 g, 320 g (model dependent) |
| Readability | 0.0001g |
| Calibration | Internal |
| Draft protection | Doors |
| Data output port | USB, RS-232 25-pin, Bluetooth®, RS-232 9-pin, including PS/2 port |
| Warranty | 2 Years |
| Model | 324S | 324P | 224S | 124S |
| Scale interval (d) | 0.1 mg | 0.1 | 0.2 | 0.5 mg | 0.1 mg | 0.1 mg |
| Maximum capacity (Max) | 320 g | 80 | 160 | 320 g | 220 g | 120 g |
| Repeatability at 5 % load | ||||
| Standard deviation of the load values, tolerance | 0.08 mg | 0.08 mg | 0.07 mg | 0.1 mg |
| Standard deviation of the load values, typical value | 0.04 mg | 0.04 mg | 0.05 mg | 0.05 mg |
| Repeatability near Max | ||||
| Standard deviation of the load values, tolerance | 0.1 mg | 0.1 mg | 0.07 mg | 0.1 mg |
| Standard deviation of the load values, typical value | 0.05 mg | 0.05 mg | 0.05 mg | 0.05 mg |
| Linearity Deviation | ||||
| Tolerance | 0.3 mg | 0.5 mg | 0.2 mg | 0.2 mg |
| Typical value | 0.2 mg | 0.2 mg | 0.13 mg | 0.13 mg |
| Deviation when load is off-center, positions according to OIML R76 | ||||
| Test weight | 200 g | 200 g | 100 g | 50 g |
| Tolerance | 0.3 mg | 0.4 mg | 0.2 mg | 0.2 mg |
| Typical value | 0.2 mg | 0.2 mg | 0.12 mg | 0.12 mg |
| Sensitivity drift between +10 °C and +30 °C | 1 ppm/K | 1 ppm/K | 1 ppm/K | 1 ppm/K |
| Tare maximum capacity: Less than 100 % of maximum capacity | ||||
| Accuracy class according to Directive 2014/31/EU | I | I | I | I |
| Verification scale interval (e) according to Directive 2014/31/EU | 1 mg | 1 mg | 1 mg | 1 mg |
| Minimum load (Min) according to Directive 2014/31/EU | 10 mg | 10 mg | 10 mg | 10 mg |
| Minimum weight according to USP (United States Pharmacopeia), Chap. 41 | ||||
| Optimum minimum weight | 82 mg | 82 mg | 82 mg | 82 mg |
| Typical minimum weight | 82 mg | 82 mg | 100 mg | 100 mg |
| Typical stabilization time | 1 s | 1 s | 1 s | 1 s |
| Typical measurement time | 3 s | 3 s | 3 s | 3 s |
| Recommended calibration weight | ||||
| External test load | 300 g | 300 g | 200 g | 100 g |
| Accuracy class, acc. to OIML R111-1 | E2 | E2 | E2 | E2 |
| Dimensions | ||||
| Weighing module (L × W × H)* | 425 x 240 x 373 mm | |||
| Weighing pan size | 85 x 85 mm | |||
| Weight, approx.* | 8.2/10.0 kg | |||